2 shows the performance of a typical table top circulator set at 25°C.
The horizontal (or time) scale is set at 2 minutes per division and the
vertical (or temperature) scale is set at 0.02°C per division. We
can see that the performance of the circulator is quite impressive as
the temperature variation over an 18 minute time period is well below
±0.01°C. In Figure 3 we examine one of our High Precision
Circulator Baths, and while the time scale is set to the same time interval
(2 minutes per division), the temperature scale has been zoomed in to
0.002°C per division. The temperature variation of the High Precision
Circulator Baths, however, is only slightly over ±0.001°C or
one millidegree C.
|