Failure
Analysis
In
the event that you should experience a problem with a U.S. Sensor
thermistor or probe assembly, our test facilities and experienced
personnel are available to perform failure analysis on your devices.
|
U.S.
Sensor Corp. is continuously improving our measurement equipment and training
of our test personnel. At the time of this publication our total expanded
uncertainty in temperature measurement is less than ±0.003°C
from -40 to 110°C. Uncertainty above 110°C and below -40°C
is only slightly higher.
The
total expanded uncertainty includes all factors present in the temperature
measurement process: equipment, environment, personnel, etc. We evaluate
both repeatability as well as reproducibility. Uncertainty is calculated
in accordance with NIST Tech Note 1297 (root sum squares with k=2).
 |
U.S.
Sensor Corp. maintains traceability in Resistance with a complete
set of Leeds and Northrup Standard Resistors. This set of standard
resistors has calibration data that dates back to 1966. The calibration
data is plotted in control charts and our long term stability for
these devices is between 0.002% and 0.004%. Total drift since 1966.
|
|
 |